General Description
Our Bubble CPAP is a unique system that can deliver high flows of Air or Oxygen ranging from 1 to 15 LPM.
Clinical studies have proven the usefulness of high flow of gases when warmed at body temperature and humidified up to 99 to 100% relative humidity and delivered through nasal cannula. Heat and Humidity prevents airway water loss, airway cooling, thickened secretions, nasal irritation & bleeding.
It allows delivery of breathing gases heated to body temperature & at nearly 100% relative humidity by nasal cannula without drying or cooling the airways, which was not possible in the conventional bubble humidifiers.
Most respiratory diseases of the neonate occur as a result of the immaturity of the premature neonate’s lungs. The respiratory system is underdeveloped and adequate gas exchange does not take place naturally, with this, there is a need for providing external respiratory support.
BENEFITS OF USING BUBBLE CPAP
It effectively maintains Functional Residual Capacity (FRC)
Most lung diseases that lead to respiratory failure are commonly associated with a reduced FRC. Maintaining FRC is very important to premature neonates who have a greater tendency of airway closure when FRC falls below closing volume.
IT HELPS REDUCE THE INFANT’S WORK OF BREATHING (WOB)
There is a decrease in infant’s minute volume and respiratory rate with this. Chest vibrations are caused by the pressure oscillations from the bubbling. These pressure oscillations are reverted back into the infant’s airway and may provide an alternate form of gas exchange through the principle of facilitated diffusion.
TECHNICAL SPECIFICATIONS | |
---|---|
Input flow range | 4 ~ 15 lpm |
FiO2 | 21 ~100% |
Set CPAP Pessure Range | 3 ~10 cmH2O |
Maximum Pressure Limit | 17 cmH2O at 8 lpm |
Operating Temperature Range | 18~26 ° |
Storage Temperature Range | - 10 ~ 50 °C |
Power Requirements | 220 VAC, 50/60 Hz |
Optional | Air compressor, oxygen Analyser. |
Flow meter unit with needle valves for Air & Oxygen and can be used as an option for blender. | |
Delivery system (includes Humidication Chamber, Pressure relief manifold, single heater patient circuit, Gas supply line) | |
Patient interface includes nasal prongs, nasal tubing's & infant Bonnets. |