    56-0108

    HEATER ASSEMBLY
    The heater assembly located above the centre of the support structure consists of an LED examination light with dimming function and a 650 watts heating system (Silica/ quartz /Ceramic Heater) for quick warming and even distribution of heat on the baby bed.
    The heater can be swivelled in either direction for taking X-ray and doing other procedures.

     

    AIR MODE
    Comes with Air Sensor, added security cover with exclusive feature of switching the heater in case the air temperature goes above 39 ºC irrespective of what the skin temperature is at the time of event.

     

    MANUAL MODE
    The system automatically start heating from 40% in manual mode and same can be changed through setting up to 100%.
    The system automatically converts from servo mode to manual mode in case of probe failure thus providing nonstop working of the unit.

     

    THE FIRST BREATH
    It combines virtually all the technology a caregiver may need in situation needed for resuscitation. The neonate can be supplied with a constant oxygen flow through the oxygen humidifier when connected to an oxygen cylinder/wall unit. The motorized slow suction can be used with a simple flip of a switch. A digital APGAR timer is standard, which signals when 1,5 and 10 minutes have elapsed.

     

    STANDARD COMPOSITION

    • Temperature Control Mode: Pre warm, Automatic, Servo control & manual control;
    • Advanced microprocessor control system with self check on the electronic circuitry skin/air/manual
    • Skin temperature  display can be converted from ºC to ºF
    • Instant Parameter Display: set temp, skin temp, heat power rate
    • Built-in APGAR Timer and >37°Cpreset function
    • Audible/Visual Alarms: temp deviation, sensor failure, power failure, temp override
    • Tilting Bed with Rotary Radiant Head
    • X-ray Cassette Tray under infant bed, examination light of high luminance
    • Built Suction Unit
    • Integrated Timer : 1 to 59 minute with count – up and count-down features
    Technical Specification
    Heated SystemSilica/Quartz/Ceramic Heater650 watts
    Heater Life Time10,000 hrs
    Effective Area1500 cm²
    Parameter DisplaySet Temp, Skin Temp, Air Temp. Heat Power Rate, A big display is provided
    for more information and visualization in Servo and Manual Mode.
    Temperature
    control range
    and relative
    Parameters
    Control modePre-Warm Mode, Auto, Manual Mode, Servo Mode
    TimerCount Up/ Down
    Temperature control range25°C ~ 37°C, for set temperature ≤37°C
    37°C ~ 38°C, for set temperature >37°C
    Skin temperature sensor measure range25°C~40°C
    Air temperature sensor measure range25°C~42°C
    Heat Output Indicate
    Deviation Between sensor measure
    0~100%
    temperature and control temperature≤0.5°C
    Skin Sensor Precision≤0.3°C
    Bed surface temperature Uniformity≤1°C
    Temperature Rise Time≤45min
    Mattress temperature Uniformity≤2.0°C
    Radiant box horizon angle adjust range0~±90°step-less
    AlarmSensor failure alarmAlarm with audible and visual when sensor open circuit, short circuit or disconnect, cut off power
    Temperature alarmAlarm with audible and visual when displayed temperature approach >40°C ( or above set temperature)
    Deviation alarmAfter temperature equilibrium, alarm with audible & visual when displayed temperature was 1°C above or below control temperature, cut off power if above 1°C
    Power failure alarmAlarm with audible and visual when power failure
    APGAR timeAlarm with audible when run between 50 seconds to 1 minute, 4 minutes 50 seconds to 5 minutes, 9 minutes 50 seconds to 10 minutes
    Baby BedMattress Tray Size68cm×56cm (Approx)
    Mattress tilt adjust range0~±12°
    Height AdjustableYes
    Mattress tray Max. Load10 kg (Approx)
    IV Pole Max. Load2 kg
    Shelf Max. load2 kg
    X-rayYes
    APGAR TimerYes
    Pressure Adjust
    Range (Optional)
    Yes
    Operating NoiseYes
    Space for MonitorYes
    Light SourceHigh Bright Blue/ White LEDWavelength – 420- 480 μm
    Irradiance <35μm
    CastorThe system is mounted on four 12.5 cm dia castor 2 with brakes
    Environment Temperature (It is a suggestion that please do not use outside of the allow environment.)Operate temperature range18°C~30°C
    Store temperature range-20°C~+55°C
    Environment HumidityOperate humidity range30%RH~75%RH
    Store humidity range≤93%RH
    Atmospheric PressureTransport and Store atmospheric pressure range500hPa~1060hPa
    Operate atmospheric pressure range700hPa~1060hPa
    Stream velocityAmbit environment stream velocity≤0.3m/s
    PowerAC 110 / 220 V, 50Hz / 60Hz
    Input Power≤1000VA
    Fuse5 Amp
    AccessoriesSkin Probe
    Air Probe
    IV Pole
    Mattress
    Suction Unit
    Oxygen Therapy Accessories