    56-0109

    HEATER ASSEMBLY
    The heater assembly located above the centre of the support structure consists of an integral examination light and a 650 watts heating system (Silica/ quartz /Ceramic Heater) for quick warming and even distribution of heat on the baby bed.
    The heater can be swivelled in either direction for taking X-ray and doing procedures.
    The Heater is also equipped with 15 LED Phototherapy on both side to provide light intensity effective on the baby bed.

    AIR MODE OPTIONAL
    Comes with Air Sensor, added security cover with exclusive feature of switching the heater in case the air temperature goes above 39 ºC irrespective of what the skin temperature is at the time of event.

     

    LED PHOTOTHERAPY OPTIONAL
    Comes with highly efficient and bright 12/15 LED on each side of heater. The strong Lux of 24/30 LEDs cure jaundice in baby rapidly.

     

    MANUAL MODE
    The system automatically start heating from 40% in manual mode and same can be changed through setting up to 100%. The system automatically converts from servo mode to manual mode in case of probe failure thus providing nonstop working of the unit.

     

    STANDARD COMPOSITION

    • Temperature Control Mode: Pre warm, Automatic, Servo control & manual control
    • Advanced microprocessor control system with self check on the electronic circuitry skin / air / manual
    • A big LCD is provided for more information and visualizations of the heater power in manual and servo modes.
    • Skin temperature  display can be converted from ºC to ºF
    • Instant Parameter Display: set temp, skin temp, heat power rate;
    • Built-in APGAR Timer and >37°C preset function
    • Audible/Visual Alarms: temp deviation, sensor failure, power failure, temp override
    • Tilting Bed with Rotary Radiant Head
    • X-ray Cassette Tray under infant bed, examination light of high luminance
    • Use anti-blast quartz infrared tube as radiant heating source.
    Technical Specification
    PowerAC 110 / 220 V, 50Hz / 60Hz
    Input Power≤1000VA
    Heated SystemSilica/Quartz/Ceramic Heater650 watts
    PhototherapyHeater is also equipped with highly eficient and bright 12/15 LED on both side of heater to provide light intensity effective on the baby bed
    Fuse5 Amp
    Light SourceHigh Bright Blue/White LED's
    Wavelength450- 470μm
    Irradance< 35 μw/cm/nm.
    TimerCount Up/ Down
    Instant Parameter
    Display
    Set temp, skin temp, heat power rate. A big display is provided for more information and
    visualization in servo and manual mode
    Temperature Control ModePre - warm, auto, manual mode, baby mode
    Temperature control range
    and relative specification
    Control modePre-warm mode, auto, manual mode, Servo mode
    Temperature control range25°C ~ 37°C, for set temperature ≤37°C
    37°C ~ 38°C, for set temperature >37°C
    Skin temperature sensor measure range25°C~40°C
    Air temperature sensor measure range25°C~42°C
    Deviation between sensor measure
    temperature and control temperature.
    ≤0.5°C
    Skin Sensor Precision≤0.3°C
    Bed surface temperature Uniformity≤1°C
    Temperature Rise Time≤45min
    Baby bed InclinationTiltable ±12 deg
    Environment Temperature (It is a suggestion that please do not use
    outside of the allow environment.)
    Operate Temperature Range18°C~30°C
    Store Temperature Range-20°C~+55°C
    AlarmSensor failure alarmAlarm with audible and visual when sensor open circuit, short circuit or disconnect, cut off power
    Temperature alarmAlarm with audible and visual when displayed temperature approach above the set temp.
    Deviation AlarmAfter temperature equilibrium, alarm with audible and visual when displayed temperature was 1°C above or below control temperature, cut off power if above 1°C
    Environment
    Humidity
    Operate humidity range30%RH~75%RH
    Store humidity range≤93%RH
    Atomspheric PressureTransport and Store atmospheric
    pressure range
    500hPa~1060hPa
    Operate atmospheric pressure range700hPa~1060hPa
    Stream velocityAmbient environment stream velocity≤0.3m/s
    Mattress Temp. Uniformity.≤2.0°C
    Temp. rise time
    Heat output indicate
    ≤45 min
    0~100%
    Radiant box horizon angle adjust range0~±90°step-less
    Mattress Tray Size68cm×56cm (Approx)
    Mattress tilt adjust range0~±12° step-less
    Mattress Tray Max. load10 kg (Approx)
    Transfusion pole Max. load2 kg
    Shelf Max. load2 kg
    X-rayYes
    APGAR TimerYes
    Temperature deviation alarm+1.0°C
    Over temp. alarm≤38.0°C( or above set temp.)
    Pressure adjust rangeYes
    Operating noiseYes
    IV PoleYes
    Space for MonitorYes
    Light SourceHigh Bright Blue/ White LEDWavelength – 420- 480 μm
    Irradiance <35μm
    StorageThe unit comes fitted with slides out for keeping consumables
    CasterThe system is fitted with 4 high quality casters having brakes in front two casters.
    Optional : Antistatic caster
    Accessories StandardSkin Probe / Air Probe, Mattress, Casters
    OptionalSuction Unit, Resuscitation Unit, Blender, Oxygen Cylinder/Flow Meter