    56-0110

    HEATER ASSEMBLY
    The heater assembly located above the centre of the support structure consists of an LED examination light with dimming function and a 650 watts heating system (Silica/ quartz /Ceramic Heater) for quick warming and even distribution of heat on the baby bed.
    The heater can be swivelled in either direction for taking X-ray and doing other procedures.

    AIR MODE
    Comes with Air Sensor, added security cover with exclusive feature of switching the heater in case the air temperature goes above 39 ºC irrespective of what the skin temperature is at the time of event.

    LED PHOTOTHERAPY OPTIONAL
    Comes with highly efficient and bright 12/15 LED on each side of heater. The strong Lux of 24/30 LEDs cure jaundice in baby rapidly.

    MANUAL MODE
    The system automatically start heating from 40% in manual mode and same can be changed through setting up to 100%.

    The system automatically converts from servo mode to manual mode in case of probe failure thus providing nonstop working of the unit.

    THE FIRST BREATH
    The open care system combine virtually all the technology a caregiver may need in situation needed for resuscitation. The neonate can be supplied with a constant oxygen flow through the oxygen humidifier when connected to an oxygen cylinder/wall unit. The motorized slow suction can be used with a simple flip of a switch. A digital APGAR timer is standard, which signals when 1,5 and 10 minutes have elapsed.

    STANDARD COMPOSITION

    • Temperature Control Mode: Pre warm, Automatic, Servo control & manual control;
    • Advanced microprocessor control system with self check on the electronic circuitry skin/air/manual
    • Skin temperature  display can be converted from ºC to ºF
    • Instant Parameter Display: set temp, skin temp, heat power rate;
    • Built-in APGAR Timer and >37°Cpreset function
    • Audible/Visual Alarms: temp deviation, sensor failure, power failure, temp override;
    • Tilting Bed with Rotary Radiant Head;
    • X-ray Cassette Tray under infant bed, examination light of high luminance;
    • Use anti-blast quartz infrared tube as radiant heating source.
    • Built in SpO2 module to measure O2 saturation and pulse rate.
    • Integrated T-Piece resuscitator with adjustable control of PIP and PEEP
    • Built Suction Unit controlled from main panel
    • Integrated CPAP with built in blender for controlling levels of Oxygen concentration from 21% to 100% with medical air compressor.
    • Integrated Timer : 1 to 59 minute with count – up and count-down features.
    • Storage: The unit comes fitted with slides out for keeping consumables
    • Space for Monitor or accessories.
    Technical Specifications
    PowerAC 110 / 220 V, 50Hz / 60Hz
    Input Power≤1000VA
    Heated SystemSilica/Quartz/Ceramic Heater650 watts
    PhototherapyHeater is also equipped with highly efficient and bright 12/15 LED on both side of heater to
    provide light intensity effective on the baby bed
    Fuse5 Amp
    TimerCount Up/ Down
    Parameter DisplaySet temp, skin temp, Air temp. heat power rate, SpO2, Pulse Rate, A big display is
    provided for more information and visualization in servo and manual mode
    Control modePre-warm mode, auto, manual mode, Servo mode
    Temperature
    control range
    and relative
    specification
    Temperature control range25°C ~ 37°C, for set temperature ≤37°C
    37°C ~ 38°C, for set temperature >37°C
    Skin temperature sensor measure range25°C~40°C
    Air temperature sensor measure range25°C~42°C
    Deviation between sensor measure
    temperature and control temperature
    ≤0.5°C
    Skin Sensor Precision≤0.3°C
    Bed surface temperature Uniformity≤1°C
    Temperature Rise Time≤45min
    Baby bed InclinationTiltable ±12 deg
    AlarmSensor failure alarmAlarm with audible and visual when sensor open circuit,short circuit or disconnect, cut off power
    Temperature alarmAlarm with audible and visual when displayed
    temperature approach ≥40°C ( or above set temperature)
    Deviation alarmAfter temperature equilibrium, alarm with audible & visual when displayed temperature was 1°C above or below control temperature, cut off power if above 1°C
    Power failure alarmAlarm with audible and visual when power failure
    APGAR timeAlarm with audible when run between 50
    seconds to 1 minute, 4 minutes 50 seconds
    to 5 minutes, 9 minutes 50 seconds to 10 minutes
    Environment Temp.
    (It is a suggestion
    that please do not
    use outside of the
    allow environment.)
    Operate Temperature Range18°C~30°C
    Store Temperature Range-20°C~+55°C
    Environment
    Humidity
    Operate humidity range30%RH~75%RH
    Store humidity range≤93%RH
    Atmospheric
    Pressure
    Transport and Store atmospheric
    Pressure range
    500hPa~1060hPa
    Operate atmospheric pressure range700hPa~1060hPa
    Stream velocityAmbient environment stream velocity≤0.3m/s
    Mattress Temp.
    Uniformity
    ≤2.0°C
    Temp. rise time≤45 min
    Heat output indicate0~100%
    Radiant box horizon
    Angle adjust range
    0~±90°step-less
    Mattress Tray Size68cm×56cm (Approx)
    Mattress tilt
    adjust range
    0~±12° step-less
    Mattress Tray
    Max. load
    10 kg (Approx)
    Transfusion pole
    Max. load
    2 kg
    Shelf Max. load2 kg
    X-rayYes
    APGAR TimerYes
    Temp. deviation alarm±1.0°C
    Over temp. alarm>40°C ( or above set temperature)
    Pressure adjust rangeYes
    Operating noiseYes
    IV PoleYes
    Space for MonitorYes
    Light SourceHigh Bright Blue/ White LED
    Wavelength – 420- 480 μm
    Irradiance <35μm
    StorageThe unit comes fitted with slides out for keeping consumables
    CasterThe system is fitted with 4 high quality casters having brakes
    AccessoriesCastor :The system is tted with 4 high quality castors having brakes in front.
    Skin Probe, Air Probe, IV Pole , Memory Foam Mattress, Acrylic Helmet, Eye Pads