Fill the details to complete enquire now
HEATER ASSEMBLY
The heater assembly located above the centre of the support structure consists of an LED examination light with dimming function and a 650 watts heating system (Silica/ quartz /Ceramic Heater) for quick warming and even distribution of heat on the baby bed.
The heater can be swivelled in either direction for taking X-ray and doing other procedures.
AIR MODE
Comes with Air Sensor, added security cover with exclusive feature of switching the heater in case the air temperature goes above 39 ºC irrespective of what the skin temperature is at the time of event.
LED PHOTOTHERAPY OPTIONAL
Comes with highly efficient and bright 12/15 LED on each side of heater. The strong Lux of 24/30 LEDs cure jaundice in baby rapidly.
MANUAL MODE
The system automatically start heating from 40% in manual mode and same can be changed through setting up to 100%.
The system automatically converts from servo mode to manual mode in case of probe failure thus providing nonstop working of the unit.
THE FIRST BREATH
The open care system combine virtually all the technology a caregiver may need in situation needed for resuscitation. The neonate can be supplied with a constant oxygen flow through the oxygen humidifier when connected to an oxygen cylinder/wall unit. The motorized slow suction can be used with a simple flip of a switch. A digital APGAR timer is standard, which signals when 1,5 and 10 minutes have elapsed.
STANDARD COMPOSITION
Technical Specifications | ||
---|---|---|
Power | AC 110 / 220 V, 50Hz / 60Hz | |
Input Power | ≤1000VA | |
Heated System | Silica/Quartz/Ceramic Heater | 650 watts |
Phototherapy | Heater is also equipped with highly efficient and bright 12/15 LED on both side of heater to provide light intensity effective on the baby bed | |
Fuse | 5 Amp | |
Timer | Count Up/ Down | |
Parameter Display | Set temp, skin temp, Air temp. heat power rate, SpO2, Pulse Rate, A big display is provided for more information and visualization in servo and manual mode | |
Control mode | Pre-warm mode, auto, manual mode, Servo mode | |
Temperature control range and relative specification | Temperature control range | 25°C ~ 37°C, for set temperature ≤37°C |
37°C ~ 38°C, for set temperature >37°C | ||
Skin temperature sensor measure range | 25°C~40°C | |
Air temperature sensor measure range | 25°C~42°C | |
Deviation between sensor measure temperature and control temperature | ≤0.5°C | |
Skin Sensor Precision | ≤0.3°C | |
Bed surface temperature Uniformity | ≤1°C | |
Temperature Rise Time | ≤45min | |
Baby bed Inclination | Tiltable ±12 deg | |
Alarm | Sensor failure alarm | Alarm with audible and visual when sensor open circuit,short circuit or disconnect, cut off power |
Temperature alarm | Alarm with audible and visual when displayed temperature approach ≥40°C ( or above set temperature) | |
Deviation alarm | After temperature equilibrium, alarm with audible & visual when displayed temperature was 1°C above or below control temperature, cut off power if above 1°C | |
Power failure alarm | Alarm with audible and visual when power failure | |
APGAR time | Alarm with audible when run between 50 seconds to 1 minute, 4 minutes 50 seconds to 5 minutes, 9 minutes 50 seconds to 10 minutes | |
Environment Temp. (It is a suggestion that please do not use outside of the allow environment.) | Operate Temperature Range | 18°C~30°C |
Store Temperature Range | -20°C~+55°C | |
Environment Humidity | Operate humidity range | 30%RH~75%RH |
Store humidity range | ≤93%RH | |
Atmospheric Pressure | Transport and Store atmospheric Pressure range | 500hPa~1060hPa |
Operate atmospheric pressure range | 700hPa~1060hPa | |
Stream velocity | Ambient environment stream velocity | ≤0.3m/s |
Mattress Temp. Uniformity | ≤2.0°C | |
Temp. rise time | ≤45 min | |
Heat output indicate | 0~100% | |
Radiant box horizon Angle adjust range | 0~±90°step-less | |
Mattress Tray Size | 68cm×56cm (Approx) | |
Mattress tilt adjust range | 0~±12° step-less | |
Mattress Tray Max. load | 10 kg (Approx) | |
Transfusion pole Max. load | 2 kg | |
Shelf Max. load | 2 kg | |
X-ray | Yes | |
APGAR Timer | Yes | |
Temp. deviation alarm | ±1.0°C | |
Over temp. alarm | >40°C ( or above set temperature) | |
Pressure adjust range | Yes | |
Operating noise | Yes | |
IV Pole | Yes | |
Space for Monitor | Yes | |
Light Source | High Bright Blue/ White LED | Wavelength – 420- 480 μm |
Irradiance <35μm | ||
Storage | The unit comes fitted with slides out for keeping consumables | |
Caster | The system is fitted with 4 high quality casters having brakes | |
Accessories | Castor :The system is tted with 4 high quality castors having brakes in front. Skin Probe, Air Probe, IV Pole , Memory Foam Mattress, Acrylic Helmet, Eye Pads |